See the table below for a description of A1CNow Self Check test system operating and error codes. (OR = Out of Range; QC = Quality Control; E = Analyzer Error

Message Description and Resolution
OR1 The blood sample may have too little hemoglobin for the test to work properly, or you added too little blood.
Call customer service.
OR2 The blood sample may have too much hemoglobin for the test to work properly, or you added too much blood.
Call customer service.
OR3 The blood sample may have too little hemoglobin A1C for the test to work properly, or you added too little blood.
Call customer service.
OR4 The blood sample may have too much hemoglobin A1C for the test to work properly, or you added too much blood.
Call customer service.
OR5 The analyzer temperature is below 64°F (18°C). The test must be repeated with a new cartridge at room temperature 64-77°F (18-25°C)
OR6 The analyzer temperature is above 77°F (18°C). The test must be repeated with a new cartridge at room temperature 64-77°F (18-25°C)
<4.0 The % A1C is less than 4%.
Call customer service.
>13.0 The % A1C is greater than 13.0%.
Call your healthcare professional.
QC2 Occurs when you insert a cartridge that already has sample added to it. Do not remove and reinsert a cartridge after adding sample.
QC6 Sample was added to cartridge before "SMPL" display. This counts down one test on the analyzer. Remove and discard cartridge.
To avoid this error, do not add sample until the "WAIT" prompt clears and "SMPL" appears.
QC7 The cartridge remained in the analyzer without sample addition for 2 minutes after "SMPL" prompt. This counts down one test on the analyzer. Discard the test cartridge and insert a fresh one when you are ready to dispense the shaker.
QC 30 - 33 The analyzer was unable to obtain a valid initial reading. Be sure to remove the shaker within one second after dispensing it into the sample port, and do not disturb the analyzer while the test is running.
QC 50 - 51
QC 55 - 56
Insufficient sample was delivered to the test cartridge.
To avoid this error be sure to fully insert the blood collector into the shaker and shake immediately.
All other QC codes The quality control checks inside the analyzer did not pass. The test will need to be repeated with another kit.
Call customer service.
E The analyzer is not working.
Call customer service.

Customer Service: 1-877-870-5610

A1CNow SELFCHECK (Monitor w/ 2 tests)
A1CNow SELFCHECK (Monitor w/ 2 tests)
Price:$43.52/strip

More Info
A1CNow+ Meter w/10 kits Professional Use Only
A1CNow+ Meter w/10 kits Professional Use Only
Price:$142.95/strip

More Info
A1CNow+ Meter w/20 kits Professional Use Only
A1CNow+ Meter w/20 kits Professional Use Only
Price:$239.99/strip

More Info
CardioChek ST Analyzer
CardioChek ST Analyzer
Price:$251.55/meter

More Info
Glucose CardioChek Panels - 25 strips
Glucose CardioChek Panels - 25 strips
Price:$24.99/strip

More Info