ChekMate Quality Control kit contains 4 specially-configured check strips and 2 MEMo Chips� which allow the user to test up to two devices simultaneously to verify performance. Two level dry strip quality system to verify CardioChek PA/Plus device.
For use with Silver CardioChek PA (version 2.6 & up)
Analyzer sold separately.
A two level dry strip quality control system designed to verify the performance of the Silver CardioChek PA (version 2.6 & up) device.
The CardioChek ChekMate™ is a two level, dry strip quality control system designed to verify the performance of the CardioChek® PA device. CardioChek Chekmate is to be used in conjunction with liquid PTS Panels® Quality Controls.
Device Quality Control Strips To check device operation, it is recommended that ChekMate Quality Control strips be run once each day of device use. Both Level 1 and Level 2 ChekMate Quality Control strips will report four (4) results. These results (C1 RED, C2 RED, C3 RED, C1 GREEN) correspond to a specific optical reading for each chemistry test “window”.
To check the performance of the entire CardioChek test system: CardioChek PA device, MEMo Chip®, and PTS Panels test strips; it is recommended that the liquid PTS Panels Quality Controls be run whenever a new lot of PTS Panels test strips are received or as required by local accrediting and regulatory bodies. Please refer to the liquid PTS Panels Quality Controls package inserts for instructions on how to run liquid quality controls.
Store in a cool dry place at room temperature of 68-80°F (20-27°C) and 20-80% relative humidity (RH). Always store in the original container when not in use to protect from light and from accidental scratching. Do not freeze.
Results are displayed as the following readings: C1 RED; C2 RED; C3; RED; C1 GREEN. These correspond to each chemistry test “window”. All results for the ChekMate Quality Control strips should be within the ranges specified on the ChekMate Quality Control Result sheet or listed at www.cardiochek.com. Users may also track their own results and define a narrower acceptance limit based upon the trend for each device as test results accumulate. If test results are outside the specified range, re-test. Should unexpected results be observed, suspect control strip damage and replace immediately.
Must call
. Be sure to mention Offer number when you call.*Excludes Shipping
Must call
. Be sure to mention Offer number when you call.*Excludes Shipping