![]() |
| |||||||
Accuracy of the CardioChek™ P·A and the Cholestech LDX® Systems Compared to Cholesterol Reference Method Laboratory Network (CRMLN) Determinations of Lipids
Abstract
Both the CardioChek P·A and the Cholestech LDX Systems measure lipids (cholesterol, HDL cholesterol and
triglycerides) in whole blood. Venous whole blood and serum samples were collected from 43 subjects in
the same blood draw. The serum was sent to a CRMLN network laboratory for cholesterol, HDL and
triglycerides analysis. The whole blood was run on both the CardioChek PA and the Cholestech LDX. Accuracy
was measured as bias between each method and the CRMLN laboratory results. The CardioChek
P·A produced
cholesterol, HDL and calculated LDL cholesterol results that had a very small bias to CRMLN laboratory
results. The Cholestech LDX System results for cholesterol and calculated LDL cholesterol showed a significant
bias when compared to the CRMLN laboratory results, as did Cholestech LDX HDL results at the second HDL
clinical cutpoint (60mg/dL HDL).
|
Introduction The CardioChek P·A (Polymer Technology Systems, Inc., Indianapolis, IN) is a handheld clinical analyzer capable of running a variety of tests on whole blood using disposable test strips. The Cholestech LDX (Cholestech Corporation, Hayward, CA) is a clinical analyzer capable of running whole blood samples using a disposable cassette. The objective of this study was to compare the test results for cholesterol, HDL cholesterol, triglycerides and calculated LDL cholesterol from both the CardioChek P·A system and the Cholestech LDX system to test results from a CRMLN reference laboratory.
|
Method The PTS PANELS Lipid Panel Test Strips and the Cholestech Lipid Profile cassettes measure cholesterol, HDL cholesterol and triglycerides and calculate LDL cholesterol. The K2EDTA whole blood was tested on both a CardioChek P·A system and a Cholestech LDX system. The PTS PANELS Lipid Panel Test Strips were used with the CardioChek P·A and the Cholestech Lipid Profile cassettes were used with the LDX. The tests were performed according to each manufacturer's instructions. The serum was sent to a CRMLN reference laboratory for cholesterol, HDL and triglycerides analysis. LDL results were calculated. The results from each of the CardioChek P·A and Cholestech LDX systems were compared to the CRMLN reference laboratory results using least squares linear regression. The regression results are listed, by test, in Table 1. Accuracy was measured as bias between each method and the CRMLN laboratory results. Bias was assessed two ways: by calculating the mean bias for each method to the CRMLN results for that test and by calculating the bias to the CRMLN results at the clinical cutpoints using each method's regression equation. The clinical cutpoints used are 200 and 240 mg/dL (cholesterol); 40 and 60 mg/dL (HDL); 150 and 200 mg/dL (triglycerides) and 130 and 160 mg/dL (LDL). |
Table 1: Accuracy of Two Analyzers Compared to the CRMLN Reference Laboratory
| Test System | Slope | Intercept | Mean Bias | Bias at Cutpoints | Range of Samples |
| Cholesterol (n=43) | 112-294 mg/dL | ||||
| CardioChek P·A vs. CRMLN | 0.976 | 8.46 | 1.95% | 1.83%, 1.13% | |
| Cholestech LDX vs. CRMLN | 1.187 | -12.32 | 12.36% | 12.54%, 13.57% | |
| HDL Cholesterol (n=42) | 30-85 mg/dL | ||||
| CardioChek P·A vs. CRMLN | 0.912 | 2.65 | 3.32% | 2.18%, 4.38% | |
| Cholestech LDX vs. CRMLN | 1.358 | -16.00 | 2.89% | 4.20%, 9.13% | |
| Triglycerides (n=39) | 46-411 mg/dL | ||||
| CardioChek P·A vs. CRMLN | 1.151 | -33.97 | 10.62% | 7.55%, 1.89% | |
| Cholestech LDX vs. CRMLN | 1.123 | -2.07 | 6.04% | 10.92%, 11.27% | |
| LDL Cholesterol (n=38) | 45-209 mg/dL | ||||
| CardioChek P·A vs. CRMLN | 0.804 | 29.41 | 5.84% | 3.02%, 1.22% | |
| Cholestech LDX vs. CRMLN | 1.181 | -1.13 | 17.15% | 17.23%, 17.39% |
|
Results The Cholestech LDX cholesterol and LDL biases to the CRMLN results were significantly higher than those from the CardioChek P·A. The HDL cholesterol mean biases compared to the CRMLN results were equivalent for the Cholestech LDX and the CardioChek P·A. However, the HDL cholesterol bias for the Cholestech LDX was significantly higher than for the CardioChek P·A at the second clinical cutpoint (60mg/dL HDL). Discussion Even though there are no National Cholesterol |
Education Program (NCEP) performance guidelines for whole blood lipid analyzers, the CardioChek P·A met the NCEP serum performance bias guideline at both clinical cutpoints for both cholesterol and HDL cholesterol. The Cholestech LDX did not meet the bias guideline at the either of the clinical cutpoints for cholesterol. The Cholestech LDX HDL met the bias guideline at the first clinical cutpoint, but failed to meet the bias guideline for HDL at the second clinical cutpoint. The CardioChek P·A system compares more closely with the CRMLN reference laboratory than does the Cholestech LDX for cholesterol, HDL cholesterol and LDL cholesterol. The overall conclusion is that the CardioChek P·A Lipid Panel test results were more accurate for cholesterol, HDL cholesterol and LDL cholesterol than the Cholestech LDX Acknowledgment |
References:
CardioChek is a trademark
of Polymer Technology Systems, Inc.
Cholestech is a registered trademark of Cholestech Corporation.

* Same
day and next day shipments are normally the case with the exception of any
out-of-stock items.
** Lifetime
member discounts are subject to member
terms & conditions. Also, use of this site & products sales that
result from this site are subject to our company's
policies & disclaimers