|
Introduction
In 1990, the Centers for Disease Control and Prevention
(CDC) established the Cholesterol Reference Method
Laboratory Network (CRMLN) to provide reference
laboratory services to manufacturers so that lipid tests
can be confirmed as traceable to the National Reference
System for Cholesterol (NRS/Chol). There are currently
four CRMLN laboratories in the United States. The
Division of Laboratory Sciences of the CDC maintains
the manufacturer's certification program for lipids. As
the CRMLN laboratories are responsible for confirming
lipid traceability to the NRS/Chol, the CRMLN
laboratory results are the most appropriate results for
assessing accuracy of a laboratory method. Both the
cholesterol and HDL cholesterol tests are standardized
through the CRMLN. There is currently no
standardization program for triglycerides.
The CardioChek P·A (Polymer Technology Systems,
Inc., Indianapolis, IN) is a handheld clinical analyzer
capable of running a variety of tests on whole blood
using disposable test strips.
The Cholestech LDX (Cholestech Corporation,
Hayward, CA) is a clinical analyzer capable of running
whole blood samples using a disposable cassette.
The objective of this study was to compare the test
results for cholesterol, HDL cholesterol, triglycerides
and calculated LDL cholesterol from both the
CardioChek P·A system and the Cholestech LDX
system to test results from a CRMLN reference
laboratory.
|
Method
Forty-three (43) subjects participated in this study.
Venous blood was drawn in both a K2EDTA tube and in
a serum tube.
The PTS PANELS Lipid Panel Test Strips and the Cholestech Lipid Profile cassettes measure cholesterol,
HDL cholesterol and triglycerides and calculate LDL
cholesterol. The K2EDTA whole blood was tested on
both a CardioChek P·A system and a Cholestech LDX
system. The PTS PANELS Lipid Panel Test Strips were
used with the CardioChek P·A and the Cholestech Lipid
Profile cassettes were used with the LDX. The tests were
performed according to each manufacturer's
instructions.
The serum was sent to a CRMLN reference laboratory
for cholesterol, HDL and triglycerides analysis. LDL
results were calculated.
The results from each of the CardioChek
P·A and
Cholestech LDX systems were compared to the CRMLN
reference laboratory results using least squares linear
regression. The regression results are listed, by test, in Table 1. Accuracy was measured as bias between each
method and the CRMLN laboratory results. Bias was
assessed two ways: by calculating the mean bias for each
method to the CRMLN results for that test and by
calculating the bias to the CRMLN results at the clinical
cutpoints using each method's regression equation.
The clinical cutpoints used are 200 and 240 mg/dL
(cholesterol); 40 and 60 mg/dL (HDL); 150 and 200 mg/dL (triglycerides) and 130 and 160
mg/dL (LDL).
|