Accuracy/Method
Correlation
Reference Methods:
Cholesterol: Abell-Kendall
traceable method and PTS PANELS Cholesterol Test Strips
HDL Cholesterol: Abell-Kendall traceable method and PTS PANELS HDL
Cholesterol Test Strips
Triglycerides: CRMLN reference method and PTS PANELS Triglycerides Test
Strips
Samples for Cholesterol and HDL cholesterol were run by the Abell-Kendall method by a certified Cholesterol Reference Method Laboratory Network (CRMLN) lab. Triglycerides were also run at the same CRMLN lab using an enzymatic assay. Method standardization studies were also performed for PTS test methods in comparison to CDC reference methods at a CRMLN laboratory. These studies are done to validate appropriate calibration of a manufacturers' lipid testing method.2
Additionally, Lipid Panel Test Strips were run by laboratory professionals and the results were compared to results from PTS PANELS Test Strips run on a PTS BioScanner 2000 instrument. Results from both methods were then compared using least squares linear regression.
RESULTS
| A | B |
![]() |
![]() |
|
C |
|
![]() |
Fig. 1 Linear regression plots for PTS Methods and Reference Methods. (A) Total Cholesterol, Equation for line: PTS Cholesterol = -1.83 + 1.01 (Abell-Kendall cholesterol); r = 0.91, N = 125. (B) Triglycerides, Equation for line: PTS Triglycerides = 2.8 + 0.97(CRMLN Laboratory Trig. Method); r = 0.97, N = 111 (C) HDL Cholesterol, Equation for line: PTS HDL Cholesterol = -4.1 + 1.1 (Abell-Kendall HDL Chol.); r = 0.89, N = 87. |
|
Table 1. Lipid Panel Method Comparison | |||
| Lipid Panel - Cholesterol | Lipid Panel - HDL Cholesterol | Lipid Panel - Triglycerides | |
| No. of Samples | 95 | 95 | 95 |
| Correlation Coefficient (r) | 0.90 | 0.90 | 0.96 |
| Bias | at 200 mg/dL, +0.05% | at 40 mg/dL, -3.25% | at 150 mg/dL, +1.32% |
| Bias | at 240 mg/dL, -3.13% | At 60 mg/dL, +4.17% | at 200 mg/dL, +3.24% |
PTS PANELS Cholesterol (r=0.91, Fig. 1A), Triglycerides (r=0.97, Fig. 1B), and HDL Cholesterol (r=0.89, Fig. 1C), correlated well to the Abell-Kendall traceable methods for total cholesterol and HDL cholesterol, as well as to the CRMLN triglycerides laboratory method. In comparison of the Lipid Panel Test Strips against the single test method, Cholesterol (r=0.91, Table 1), Triglycerides (r=0.96, Table 1), and HDL Cholesterol (r=0.90, Table 1) all correlated well. Overall, a small bias was observed, between 3 to 4% (Table 1) at clinical decision points for cholesterol, HDL cholesterol, and triglycerides testing.
Method standardization studies demonstrated that the Lipid Panel Test Strips are standardized to methods traceable to CRMLN laboratory methods. Figure 2 demonstrates the very high level of agreement of the PTS reference method to the CRMLN values for cholesterol (r=0.989), HDL cholesterol (r=0.989), and triglycerides (r=0.994).
| A | B |
![]() |
![]() |
|
C |
|
![]() |
Fig.
2 Method standardization comparison. (A). Total Cholesterol, Equation for line: PTS Reference Cholesterol 3.1 + 0.924(CRMLN Cholesterol); r = 0.989, N = 43. (B) HDL Cholesterol, Equation for line: PTS Reference HDL Cholesterol = 3.1 + 0.924(CRMLN HDL Cholesterol); r = 0.989, N =43. (C) Triglycerides, Equation for line: PTS Reference Triglycerides 0.098 + 0.955(CRMLN Triglyceride); r = 0.994, N = 44. |

* Same
day and next day shipments are normally the case with the exception of any
out-of-stock items.
** Lifetime
member discounts are subject to member
terms & conditions. Also, use of this site & products sales that
result from this site are subject to our company's
policies & disclaimers