Error Code |
Explanation |
How
to Correct |
| Er1 |
Inflation error
- When the pressure
does not exceed 12 mmHg
within the set time
after the start of
inflation
- When the inflation
does not reach the set
cuff Pressure within the
specified time after the
start of inflation
|
- Confirm that the air
tube connecting the cuff
and the main unit is
connected securely.
- Confirm that the air
flow in the air tube
connecting the cuff and
the main unit isn't
being restricted.
- Confirm that the
cuff is wrapped
correctly
(refer to page 14 and 15 in
the instructions)
- Check Bladder for
leaks and, if neccessary,
replace the bladder with
a new one.
|
|
Er2 |
Deflation Error
- When the deflation
speed is too fast during
the measurement
- When the deflation
speed is too slow during
the measurement
- When the measurement
does not finish within
the specified time after
starting the
measurement.
|
| Er3 |
Overpressure Error
- The cuff pressure
exceeded 299 mmHg
|
- Confirm that air
flow in the air tube
connecting the cuff and
the main unit isn't
being restricted.
|
| Er4 |
Insufficient Inflation
error
- Blood Pressure could
not be measured due
insufficient inflation
level.
|
- If the measurement
is made by setting the
P-SET to "AUTO", ask the
patient not to move
during the inflation.
- Confirm that the
P-SET is securely set to
"AUTO". Turn the knob
counterclockwise as far
as it goes until you can
hear a click sound.
- If the measurement
is made by manual
inflation level setting,
set the value to 30 to
40 mmHg higher.
|
| Er5 |
Indeterminable Blood
Pressure error
- Blood Pressure could
not be measured even
when the cuff pressure
reached the specified
pressured.
|
|
| Er6 |
Low pulse level error
- Pulse wave was too
small.
|
|
| Er7 |
Blood Pressure error
- Relationship between
systolic and diastolic
pressures was abnormal
|
- Ask the patient not
to move during the
measurement
- Check the patient
for arrhythmia.
|
| Er8 |
Pulse rate error
- Pulse rate did not
stay within the range of
30 to 199 beats/min
|
| Er9 |
Device Error
|
- Contact Omron's
repair department
toll-free at
1-877-216-1336
|