Error Code Explanation How to Correct
Er1 Inflation error
  • When the pressure does not exceed 12 mmHg within the set time after the start of inflation
  • When the inflation does not reach the set cuff Pressure within the specified time after the start of inflation
  • Confirm that the air tube connecting the cuff and the main unit is connected securely.
  • Confirm that the air flow in the air tube connecting the cuff and the main unit isn't being restricted.
  • Confirm that the cuff is wrapped correctly (refer to page 14 and 15 in the instructions)
  • Check Bladder for leaks and, if neccessary, replace the bladder with a new one. 
Er2 Deflation Error
  • When the deflation speed is too fast during the measurement
  • When the deflation speed is too slow during the measurement
  • When the measurement does not finish within the specified time after starting the measurement.
Er3 Overpressure Error
  • The cuff pressure exceeded 299 mmHg
  • Confirm that air flow in the air tube connecting the cuff and the main unit isn't being restricted.
Er4 Insufficient Inflation error
  • Blood Pressure could not be measured due insufficient inflation level.
  • If the measurement is made by setting the P-SET to "AUTO", ask the patient not to move during the inflation.
  • Confirm that the P-SET is securely set to "AUTO". Turn the knob counterclockwise as far as it goes until you can hear a click sound.
  • If the measurement is made by manual inflation level setting, set the value to 30 to 40 mmHg higher.
Er5 Indeterminable Blood Pressure error
  • Blood Pressure could not be measured even when the cuff pressure reached the specified pressured.
Er6 Low pulse level error
  • Pulse wave was too small.
Er7 Blood Pressure error
  • Relationship between systolic and diastolic pressures was abnormal
  • Ask the patient not to move during the measurement
  • Check the patient for arrhythmia.

 

Er8 Pulse rate error
  • Pulse rate did not stay within the range of 30 to 199 beats/min
Er9 Device Error
  • Main Unit malfunction
  • Contact Omron's repair department toll-free at 1-877-216-1336


Product Home Page Features & Benefits Symptoms & Conditions
Instructions Performance Characteristics Compounds
Cross Reactivity Drug Detection Times FDA Clearance
Product Information Principle Sensitivity Treating Teens Book
Buy This Product


*   Same day and next day shipments are normally the case with the exception of any out-of-stock items.
** Lifetime member discounts are subject to member terms & conditions. Also, use of this site & products sales that result from this site are subject to our  company's policies & disclaimers
 

Click to verify BBB accreditation and to see a BBB report.  
 



Test Medical Symptoms
6633 Ashman Rd. Maria Stein, OH 45860
PH: 1-888-595-3136
PH: 1-419-586-9462
FX: 1-775-703-6393

Email Us

Copyright
2001-2009
Family-Run
 Business
Since 2001
Authorized
CardioChek Dealer
Better
Business
Bureau
At Rating
SSL Secured
Checkout
Payment Types
Accepted
Shipping
& Tracking

Send this site to a friend! (click here)