SalivaConfirm™
- Oraline - Premium Saliva Test Instructions: |
SalivaConfirm ™- ORALINE®
IV TEST PROCEDURE One-Step, Drugs of Abuse Oral Fluid Test Kit for
Marijuana, Cocaine, Methamphetamines & Opiates.
TEST PREPARATION:
1. If the test device was stored refrigerated, allow
the test device to come to room temperature prior to opening the pouch.
2. Make sure that the person being tested has had
nothing in his/her mouth for at least five (5) minutes before the oral
fluid sample is collected (including food and chewing gum).
3. The person being tested must be able to produce
and deliver a sufficient volume of oral fluid sample (about 0.4 ml) with
minimum amount of bubbles in order for the test to function properly. In
case of collection difficulty, administrator may choose to collect the
oral fluid in a disposable cup first. The oral fluid could be delivered
into the spoon by pouring or using a pipette.
NOTE: If
the person being tested is experiencing dry mouth prior to sample
collection, have the person drink a glass of water or have a piece of
candy, wait five (5) minutes then proceed with administration of the
test . Or ask the subject to deposit the oral fluid into a cup
first, then transfer the sample to the spoon as illustrated in # 3A &
3B.
Test Procedure:
|
1. Remove the test device
from the pouch. Verify that five (5) green lines are present in
the test window (thin arrow). If the green lines are not
present, a new device with 5 visible green lines should be used.
Mark the device with the subject's name or ID number (wide
arrow) and remove the cap.
2. Prepare the test subject for saliva sample
collection. Let the person accumulate sufficient saliva in the
mouth, first, for about a minute.
3 . Place the spoon end of
the device into the subject's mouth and collect a spoonful of
oral fluid. Make sure the spoon is in a horizontal position
with the handle upward as illustrated.
If pouring the sample from a cup, also ensure
that the spoon is held horizontally with the handle in an upward
position.
|
Alternatively:
|
3A. Collect oral fluid in
the small plastic cup (0.5ml – 1.0 ml).
3B. Transfer the sample into
the spoon as illustrated. |
|
4. Once the spoon is filled,
do not replace the cap until the sample shows up in the view
window. Place the device on a protected flat surface, and allow
to run for 10 minutes
5. The test results are read
between 10-12 minutes in the viewing window. Make sure the
control line (C) is clearly visible before reading and recording
the results.
|
INTERPRETATION OF TEST RESULT:
For normal negatives, the red lines will replace all
green lines as the test runs. Absence of any test line is presumptive
positive for the tested drug. The SalivaConfirm - OraLine® test has
shown effective in detecting recent drug usage. (Please refer to the
product insert for detailed description)
NOTE : If
the test is read before ten (10) minutes or after twelve (12) minutes,
the results may be different. Thus, it is most important to record the
results within 10-12 minutes. |