Technical Bulletin

BioSafe Laboratories, Inc. 
Hemoglobin A1c Test 

Dried Blood Spot Hemoglobin A1c Assay 
Performance Characteristics 

Within-Run Precision 
Within-run precision was determined by assaying dried blood spots containing various concentrations of Hemoglobin A1c. Each of the dried blood spot specimens was assayed in replicates of two in twenty different runs. The data are presented below.

Mean
(% High A1c)
Standard Deviation % Coefficient of
Variation
5.0 (DBS)
7.6 (DBS)
0.11
0.11
3.3%
2.6%

Between-Run Precision
Between-run precision was determined by a series of duplicate measurements of two different samples over a series of twenty different runs (n=20). The data are presented below.

Mean
(% High A1c)
Standard Deviation % Coefficient of
Variation
5.0 (DBS)
7.6 (DBS)
0.21
0.26
4.4%
3.5%

Accuracy
Paired EDTA whole blood and capillary dried blood spot samples containing varying concentrations of Hemoglobin A1c were analyzed. Percent Hemoglobin A1c concentrations observed for the dried blood spot specimens versus EDTA whole blood were analyzed by linear regression.

N=110
Correlation Coefficient = 0.98
Slope = 0.95
Intercept = 0.43
  Dried Blood
%Hgb A1c
Comparable Whole Blood
Methodology
Mean %HgbA1c 7.3 7.3
%HgbA1c Range 4.0% - 12.3% 4.4% - 12.8%
Standard Deviation
%Hgb A1c
2.1 2.0

Expected Values
A population of ninety-one apparently healthy persons was analyzed using the dried blood % Hgb A1c methodology. The expected values generated utilizing these values are presented below. 

  Lower Limit
% Hgb A1c
Upper Limit
% Hgb A1c
Expected Values 3.80% 5.90%

Specimen Requirements
The BioSafe Laboratories Dried Blood Spot Hemoglobin A1c Assay requires three (3) 50uL drops of capillary whole blood placed onto a specialized paper card. The blood is dried and placed into a waterproof outer envelope for shipping via regular mail to BioSafe Laboratories for testing.

Specimen
Dried blood spot Hemoglobin A1c specimens are stable for more than 28 days when stored at or below 37 degrees centigrade and for greater than 3 months when stored at -20 degrees C.

Convenient and Simple to Administer
Simple instructions guide the patient through self-collection of a capillary blood sample onto a collection card using a finger lancet. The BioSafe Hemoglobin A1c Detection System requires only four simple steps:

  1. Collection kit provided to patient.
  2. Patient applies blood sample onto card
  3. Blood sample sent to BioSafe Laboratories in provided wager.
  4. Hemoglobin A1c results returned within days.

 

Dried Blood Hemoglobin A1c: Specimen Stability Evaluation.Graphics (p.1-1)

Blood with varying levels of endogenous Hemoglobin A1c was spotted onto specialized paper cards, dried and stored in specimen return bags identical to those provided with the BioSafe Laboratories Hemoglobin A1c Collection Kit

The blood spots were subjected to a variety of harsh storage conditions over a period of 30 days. Periodic analysis was performed on the dried blood specimens, the results of which were compared to the original dried blood hemoglobin A1c value.

Storage Conditions

Blood spots were stored for alternating 12 hour periods at 50 degrees C and -20 degrees C to determine the effects of multiple freeze thaw cycles and exposure to high temperatures.Graphics (p.1-2)

Blood spots were stored in a humidity chamber at 34 degrees C to determine the effects of moisture on the specimen and packaging,

Results

No statistically significant change was observed over the length of the study regardless of storage conditions. The data is graphically presented.

 

  Graphics (p.1-3) Graphics (p.1-4)

 

 

 


Product Home Page Features & Benefits Symptoms & Conditions
Buy This Product


*   Same day and next day shipments are normally the case with the exception of any out-of-stock items.
** Lifetime member discounts are subject to member terms & conditions. Also, use of this site & products sales that result from this site are subject to our  company's policies & disclaimers
 

TMS
Privacy Policy
Click to verify BBB accreditation and to see a BBB report.  
 



Test Medical Symptoms
6633 Ashman Rd.
Maria Stein, OH 45860

PH: 1-888-595-3136
PH: 1-419-586-9462
FX: 1-775-703-6393

Email Us

Family-Run
 Business
Since 2001
Authorized
CardioChek Dealer
Better
Business
Bureau
At Rating
SSL Secured
Checkout
Payment Types
Accepted
Shipping
& Tracking

Send this site to a friend! (click here)